Features
This ppm-order zirconia oxygen analyzer enables continuous measurement by adding a built-in gas suction pump to our well-established LC-750L model―a ppm-order zirconia oxygen analyzer which employs Toray Engineering's unique zirconia sensor, and is suitable for control and monitoring of processes such as atmosphere treatment and baking, and all types of experiments and research.
Main Specifications
| Type | Portable or panel-mounted | |
|---|---|---|
| Display system | Digital | Analog |
| Measurement range | 0-10 / 100 / 1000 ppm / 100 vol%O2 | |
| Response speed | Within 90 sec (90 % response) | |
| Repeatability | Within ±0.5 % of full scale (at or above 1 % range) Within ±1 % of full scale (below 1% range) |
|
| Sampling system | Continuous suction | |
| Reference gas | 1000-2000 mL/min | |
| Reference gas | Air | |
| Recorder output | DC 0-10 V 4-20 mA |
|
| Alarm output | Upper/Lower limit alarm | |
| Ambient temperature | 0-40 °C | |
| Warmup time | Approx. 20 min | |
| Dimensions | 210(W) × 288(H) × 327(D) mm | |
| Weight | Approx. 10 kg | |
| Panel cut dimensions | 225(W) × 269(H) | |
| Gas connection | Rc1/4 (Female) | |
| Power supply | AC 100 V 50/60 Hz | |
| Max. power consumption | 200 VA | |
Specifications are subject to change due to product improvements.