DFL-1/DFL-5 online float-type density meters
For ultra-high-resolution/high-resolution fluid measurement
DFL-1 ultra-high-resolution fluid measurement densitometer
- The online densitometer for high-sensitivity measurement of fluid density.
Ultra-high-resolution density measurement with a maximum sensitivity of 0.00005 g/cm³.
The density measurement determines the quality and purity of measured fluids and the degree to which reactions have progressed. - Measurement is performed by the converting float buoyancy into air pressure, requiring no electricity.
- The function of automatic compensating performs for the changing density caused by increases or decreases in measured fluids temperature.
- Available to measuring pressurized fluids.
Wet measurement method available to measuring at pressures of up to 1.961 MPa (20 kgf/cm²). - Simple balance scale design and structure for easy maintenance.
- Track records have been over 2,100 units since 1970.
DFL-1 specifications
Measurement range | Span can be set as desired between 0.01 and 0.1 g/cm³ |
---|---|
Operating temperature | Max.100℃ |
Operating pressure | Max.1.961MPa (20 kgf/cm²) |
Correction range | Within ±20°C of measurement fluid reference temperature |
Output signal | 20 to 100 kPa (can output 4 to 20 mADC when using pneumatic-electro converter) |
Measurement fluid flow rate | 5 L/min or less |
Measured fluid viscosity | 2 to 3 mPa・s (cp) or less |
Installation method | Installation method |
Measurement precision | Within ±0.5% of span |
Unit weight | Approx. 45 kg (without float chamber)/approx. 65 kg (with float chamber) |
Dimensions | Individual design (without float chamber)/400 mm (W) x 220 mm (D) x 1,300 mm (H) (with float chamber) |
DFL-5 high-resolution fluid measurement densitometer
- The online densitometer for high-sensitivity measurement of fluid density.
High-resolution density measurement with a maximum sensitivity of 0.00025 g/cm³.
The density measurement determines the quality and purity of measured fluids and the degree to which reactions have progressed. - Measurement is performed by the converting float buoyancy into air pressure, requiring no electricity.
- The function of automatic compensating performs for the changing density caused by increases or decreases in measured fluids temperature.
- More compact and lower price than the DFL-1.
- Simple design and structure for easy maintenance.
- Track records have been over 2,100 units since 1970.
DFL-5 specifications
Measurement range | Fixed span selectable from following options: 0.05, 0.1, 0.25, 0.5 g/cm³ |
---|---|
Operating temperature | Max.100℃ |
Operating pressure | Atmospheric pressure |
Correction range | Within ±20°C of measurement fluid reference temperature |
Output signal | 20 to 100 kPa (can output 4 to 20 mADC when using pneumatic-electro converter) |
Measurement fluid Flow rate | 5 L/min or less |
Measured fluid viscosity | 2 to 3 mPa・s (cp) or less |
Installation method | Installation method |
Measurement precision | Within ±0.5% of span |
Unit weight | Approx. 10 kg (without float chamber)/approx. 30 kg (with float chamber) |
Dimensions | Individual design (without float chamber)/245 mm (W) x 140 mm (D) x 1,330 mm (H) (with float chamber) |