Oxygen AnalyzersOxygen Analyzers for Glove box
LC-450(A), LD-450/SD sensor
Model: LC-450A

- Measurement across a wide range of concentrations
Capable of displaying a wide range of concentrations, from ppm level to 100% - Improved sensor durability
Special coating used on oxygen sensor for greater durability - Wide AC power range AC100~240VAC(LC-450A)
* DC 24 V power used for LC-450D - Compact and easy to install
Shorter depth than conventional devices. Save space. - Applicable Standards
Safety:CE, UKCA- EN 61010-1
EMC: CE-EN61326-1
RoHS Directive: EN 63000
LC-450A specifications
| Measurement principle | Zirconia concentration cell type | ||
|---|---|---|---|
| Type | Desktop type/panel mounted type | ||
| Display system | 4-digit digital display (concentration display) | ||
| Measurement range | 0-1 / 10 / 100 / 1000ppm / 1 / 10 / 100vol%O2 10-20~10-0atm Automatic range | ||
| Repeatability | Concentrations in the range ±1% of full scale and 0-1 ppm, atm are out of warranty. | ||
| Sampling method | Continuous suction with built-in pump | ||
| Sample gas requirements | Pressure | Pressurization of 30 KPa or less | |
| Flow rate | 1,000~2,000mL/min | ||
| Humidity | Dew point below ambient temperature | ||
| Temperature | 0 to 50°C | ||
| Others | Must not contain corrosive gas (fluorine gas, chlorine gas, sulfuric gas, etc.), gas containing organic silicone, flammable substances (including gas), oil or moisture | ||
| Reference gas | Atmosphere | ||
| Recorder output | Oxygen concentration output | Switch between 4-20 mA or 0-20 mA (0-10 V is optional); insulated; load resistance of 550 Ω or less Recorder output range taken as full scale and oxygen concentration given in linear output |
|
| Output range | 0-100%, 0-1000ppm, 0-100ppm, 0-10ppm Automatic/fixed range | ||
| Transmission output | RS-232C | ||
| Alarm output | FAIL (device error) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when device error occurs (contact point operation adjustable) |
| ALM (concentration alarm) | 2 points (H/L) | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when alarm occurs (contact point operation adjustable) |
|
| RDY (warmup signal) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
|
|
| MRK (range marker) output | 2 points | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less | |
| Warmup time | 5 minutes or less | ||
| Dimensions | Body | 210 mm (W) ×134 mm (H)×300 mm (D) (excluding projecting parts) | |
| Weight | Approx. 6 kg | ||
| Gas connection | INLET | Rc1/4 | |
| BYPASS OUT | Rc1/4 | ||
| OUTLET | Rc1/4 | ||
| Power | 100 to 240 V AC, 50 V A (normal conditions: 35 V A) 50 Hz/60 Hz | ||
Model: LD-450/SD sensor

- Flexible setup layout. The oxygen sensor and low-profile converter are separated, greatly relaxing limitations on where the device can be installed.
- Space saving. The newly developed low-profile wall mounted converter eliminates concerns about the device’s depth.
- Ultra-high-speed response. The separated sensor provides the fastest response speed.
- Highly durable sensor. A special coating is used for even greater durability.
- Applicable Standards
CE Marking: EN 61010-1, EN 61326-1
RoHS Directive: EN 63000
Applications
Sputtering systems; equipment gas purity management; factory piping argon (Ar) welding management; experimental equipment; and testing equipment. - Sensors are compatible with NW flanges (16, 25, 40) for vacuum and screw-in units (3/8R).Select the appropriate type for use condition.
Installation example (SD sensor and LD-450 converter work in pairs.)

LD-450/SD sensor specifications
| Measurement principle | Zirconia concentration cell type | ||
|---|---|---|---|
| Type | Separate-type (converter: wall-mounted; sensor: direct insertion type) | ||
| Display system | 4-digit digital display (concentration display) | ||
| Measurement range | 0-1 / 10 / 100 / 1000ppm / 1 / 10 / 100vol%O2 10-20~10-0atm Automatic/fixed range |
||
| Repeatability | Within ±0.5% of full scale; range of 0-1% or more Within ±1% of full scale or ±1ppm, whichever is larger; range of less than 0-1% Concentrations in the 0-1 ppm, atm range are out of warranty. |
||
| Sample gas requirements |
Pressure | 10-3 to 200 KPa (absolute pressure) | |
| Flow rate | 0.5 to 5,000 mL/min or 0 mL/min (measurement in state of natural dispersion) | ||
| Humidity | Dew point below ambient temperature | ||
| Temperature | 0 to 100°C | ||
| Others | Must not contain corrosive gas (fluorine gas, chlorine gas, sulfuric gas, etc.), gas containing organic silicone, flammable substances (including gas), oil or moisture | ||
| Reference gas | Atmosphere | ||
| Recorder output | Oxygen concentration output | 4-20m (choice between this and transmission output) Insulated; load resistance of 550 Ω or less Recorder output range taken as full scale, and oxygen concentration given in linear output |
|
| Output range | 0-100%, 0-1000ppm, 0-100ppm, 0-10ppm Automatic/fixed range | ||
| Transmission output | RS-232C(choice between this and analog output) | ||
| Alarm output | FAIL (device error) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when device error occurs (contact point operation is adjustable) |
| ALM (concentration alarm) | 4 points (HH/H/L/LL) | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when alarm occurs (contact point operation is adjustable) |
|
| RDY (warmup signal) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
|
|
| MRK (range marker) output | 2 points | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.3 A or less | |
| Warmup time | 6 minutes or less | ||
| Dimensions | Converter | 185 mm (W) ×155 mm (H)×40 mm (D) (excluding projecting parts) | |
| Sensor unit | SD3/8R | Overall length 100mm(sensor ⌀ 11×37,body ⌀ 40×48.5) |
|
| SD16NW | Overall length 100mm(sensor ⌀ 11×39,body ⌀ 40×51) | ||
| SD25NW | Overall length 100mm(sensor ⌀ 11×42,body ⌀ 40×48) | ||
| SD40NW | Overall length 100mm(sensor ⌀ 11×41,body ⌀ 40×49) | ||
| Weight | Converter | Approx. 1.2 kg | |
| Sensor unit | Approx. 0.5 kg | ||
| External connection | Vacuum flange (NW16, 25, or 40) or R3/8 screw | ||
| Power | 24 V DC±10%, 60 V A (normal: 15 V A or less) | ||