Oxygen AnalyzersFor Vacuum Deposition, Sealed Furnaces, and Gloveboxes
SD/LD-450
Model: SD/LD-450
- Flexible setup layout. The oxygen sensor and low-profile converter are separated, greatly relaxing limitations on where the device can be installed.
- Space saving. The newly developed low-profile wall mounted converter eliminates concerns about the device’s depth.
- Ultra-high-speed response. The separated sensor provides the fastest response speed.
- Highly durable sensor. A special coating is used for even greater durability.
- Compliant standards. Compliant with CE marking, EN61010-1, EN61326-1, and RoHS. Sputtering systems, equipment gas purity management, factory pipe Ar weld management, experimental equipment, and testing equipment
- Sensors are compatible with NW flanges (16, 25, 40) for vacuum and screw-in units (3/8R).Select the appropriate type for use condition.
Installation example (SD sensor and LD-450 converter work in pairs.)
SD/LD-450 specifications
Measurement principle | Zirconia concentration cell type | ||
---|---|---|---|
Type | Separate-type (converter: wall-mounted; sensor: direct insertion type) | ||
Display system | 4-digit digital display (concentration display) | ||
Measurement range | 0-1 / 10 / 100 / 1000ppm / 1 / 10 / 100vol%O2 10-20~10-0atm Automatic/fixed range |
||
Repeatability | Within ±0.5% of full scale; range of 0-1% or more Within ±1% of full scale or ±1ppm, whichever is larger; range of less than 0-1% Concentrations in the 0-1 ppm, atm range are out of warranty. |
||
Sample gas requirements |
Pressure | 10-3 to 200 KPa (absolute pressure) | |
Flow rate | 0.5 to 5,000 mL/min or 0 mL/min (measurement in state of natural dispersion) | ||
Humidity | Dew point below ambient temperature | ||
Temperature | 0 to 100°C | ||
Others | Must not contain corrosive gas (fluorine gas, chlorine gas, sulfuric gas, etc.), gas containing organic silicone, flammable substances (including gas), oil or moisture | ||
Reference gas | Atmosphere | ||
Recorder output | Oxygen concentration output | 4-20m (choice between this and transmission output) Insulated; load resistance of 550 Ω or less Recorder output range taken as full scale, and oxygen concentration given in linear output |
|
Output range | 0-100%, 0-1000ppm, 0-100ppm, 0-10ppm Automatic/fixed range | ||
Transmission output | RS-232C(choice between this and analog output) | ||
Alarm output | FAIL (device error) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when device error occurs (contact point operation is adjustable) |
ALM (concentration alarm) | 4 points (HH/H/L/LL) | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less Contact point closes when alarm occurs (contact point operation is adjustable) |
|
RDY (warmup signal) | 1 point | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
|
|
MRK (range marker) output | 2 points | Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.3 A or less | |
Warmup time | 6 minutes or less | ||
Dimensions | Converter | 185 mm (W) ×155 mm (H)×40 mm (D) (excluding projecting parts) | |
Sensor unit | SD3/8R | Overall length 100mm(sensor ⌀ 11×37,body ⌀ 40×48.5) |
|
SD16NW | Overall length 100mm(sensor ⌀ 11×39,body ⌀ 40×51) | ||
SD25NW | Overall length 100mm(sensor ⌀ 11×42,body ⌀ 40×48) | ||
SD40NW | Overall length 100mm(sensor ⌀ 11×41,body ⌀ 40×49) | ||
Weight | Converter | Approx. 1.2 kg | |
Sensor unit | Approx. 0.5 kg | ||
External connection | Vacuum flange (NW16, 25, or 40) or R3/8 screw | ||
Power | 24 V DC±10%, 60 V A (normal: 15 V A or less) |