Oxygen AnalyzersFor Vacuum Deposition, Sealed Furnaces, and Gloveboxes

SD/LD-450

Model: SD/LD-450

SD/LD-450

  • Flexible setup layout. The oxygen sensor and low-profile converter are separated, greatly relaxing limitations on where the device can be installed.
  • Space saving. The newly developed low-profile wall mounted converter eliminates concerns about the device’s depth.
  • Ultra-high-speed response. The separated sensor provides the fastest response speed.
  • Highly durable sensor. A special coating is used for even greater durability.
  • Compliant standards. Compliant with CE marking, EN61010-1, EN61326-1, and RoHS. Sputtering systems, equipment gas purity management, factory pipe Ar weld management, experimental equipment, and testing equipment
  • Sensors are compatible with NW flanges (16, 25, 40) for vacuum and screw-in units (3/8R).Select the appropriate type for use condition.

Installation example (SD sensor and LD-450 converter work in pairs.)

SD/LD-450 specifications

Measurement principle Zirconia concentration cell type
Type Separate-type (converter: wall-mounted; sensor: direct insertion type)
Display system 4-digit digital display (concentration display)
Measurement range 0-1 / 10 / 100 / 1000ppm / 1 / 10 / 100vol%O2 10-20~10-0atm
Automatic/fixed range
Repeatability Within ±0.5% of full scale; range of 0-1% or more
Within ±1% of full scale or ±1ppm, whichever is larger; range of less than 0-1%
Concentrations in the 0-1 ppm, atm range are out of warranty.
Sample gas requirements
Pressure 10-3 to 200 KPa (absolute pressure)
Flow rate 0.5 to 5,000 mL/min or 0 mL/min (measurement in state of natural dispersion)
Humidity Dew point below ambient temperature
Temperature 0 to 100°C
Others Must not contain corrosive gas (fluorine gas, chlorine gas, sulfuric gas, etc.), gas containing organic silicone, flammable substances (including gas), oil or moisture
Reference gas Atmosphere
Recorder output Oxygen concentration output 4-20m (choice between this and transmission output)
Insulated; load resistance of 550 Ω or less
Recorder output range taken as full scale, and oxygen concentration given in linear output
Output range 0-100%, 0-1000ppm, 0-100ppm, 0-10ppm Automatic/fixed range
Transmission output RS-232C(choice between this and analog output)
Alarm output FAIL (device error) 1 point Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
Contact point closes when device error occurs (contact point operation is adjustable)
ALM (concentration alarm) 4 points (HH/H/L/LL) Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
Contact point closes when alarm occurs (contact point operation is adjustable)
RDY (warmup signal) 1 point Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.5 A or less
  • Minimum open/close capacity of the non-voltage contact point is 12 V or higher; does not respond to extremely small loads
MRK (range marker) output 2 points Non-voltage contact; contact capacity: 30 V DC/AC or less, 0.3 A or less
Warmup time 6 minutes or less
Dimensions Converter 185 mm (W) ×155 mm (H)×40 mm (D)  (excluding projecting parts)
Sensor unit SD3/8R Overall length 100mm(sensor ⌀ 11×37,body ⌀ 40×48.5)
SD16NW Overall length 100mm(sensor ⌀ 11×39,body ⌀ 40×51)
SD25NW Overall length 100mm(sensor ⌀ 11×42,body ⌀ 40×48)
SD40NW Overall length 100mm(sensor ⌀ 11×41,body ⌀ 40×49)
Weight Converter Approx. 1.2 kg  
Sensor unit Approx. 0.5 kg
External connection Vacuum flange (NW16, 25, or 40) or R3/8 screw
Power 24 V DC±10%, 60 V A (normal: 15 V A or less)